Microfiche records prior to 2006 have not been completely digitized and may not be available yet on PRO. If you can not find what you are looking for please submit a records request.
Permit: TC-COM-0523-01284
Permit Details
Apply Date:
05/16/2023
Applicant:
Mark Miller
Description:
This project consists of a 879 sq.ft. remodel in an existing medical building for the installation of a Biograph Horizon equipment
Permit Reviews
Permit Number - TC-COM-0523-01284
Permit Status: Complete
Permit Description: This project consists of a 879 sq.ft. remodel in an existing medical building for the installation of a Biograph Horizon equipment
Start Date | Submittal | Complete Date | Status |
---|---|---|---|
02/05/2024 | BUILDING DEFERRED SUBMITTAL REVIEW v.1 | 02/06/2024 | APPROVED |
01/24/2024 | BUILDING DEFERRED SUBMITTAL REVIEW v.1 | 01/24/2024 | APPROVED |
02/13/2024 | BUILDING DEFERRED SUBMITTAL REVIEW v.1 | 02/14/2024 | APPROVED |
06/08/2023 | COMMERCIAL REVIEW - FULL v.2 | 07/17/2023 | APPROVED |
07/17/2023 | RESTAMP - COMMERCIAL v.1 | 07/18/2023 | APPROVED |
03/19/2024 | BUILDING DEFERRED SUBMITTAL REVIEW v.1 | 03/25/2024 | APPROVED |
05/19/2023 | COMMERCIAL REVIEW - FULL v.1 | 06/05/2023 | REQUIRES RESUBMIT |
Conditions
Permit Number - TC-COM-0523-01284
Permit Status: Complete
Permit Description: This project consists of a 879 sq.ft. remodel in an existing medical building for the installation of a Biograph Horizon equipment
Date | Description | Comments |
---|---|---|
06/15/2023 | Applies to All: Deferred submittal documents must be submitted and approved prior to installation of construction elements included in deferred submittal documents. | |
06/15/2023 | Special inspections must be performed by an approved third party inspection agency at the owner's expense. Special inspection forms and reports are required to be submitted for review and approved prior to issuance of final approval documents. | |
07/27/2023 | The Contractor you have added to your permit cannot be verified. Please contact staff to have the ROC number added to the account. | |
03/18/2024 | SPECIAL INSPECTION FOR EPOXY ANCHORS IS REQUIRED. | |
06/16/2023 | SPECIAL INSPECTION FOR LEAD LINED WALL INSTALLED FOR RADIATION PROTECTION (NEW SCANNING EQUIPMENT) IS REQUIRED |
Outstanding Inspections
Permit Status: Complete
Permit Number: TC-COM-0523-01284
Permite Description: This project consists of a 879 sq.ft. remodel in an existing medical building for the installation of a Biograph Horizon equipment
Total Outstanding Inspections - 3
Last Inspection Date | Description | Inspector | Results | Comments |
---|---|---|---|---|
12/28/2023 | COT - HCBU - Building Above Ceiling | N/A | Partial Pass | Partial pending framing at duct entry from ceiling. Looked at drywall 5/8 type X nailing pattern in scanner room #105. Okay for lead line at joints and screws. 12/1/2023 New mechanical unit pad on roof per detail 201. Okay to proceed progress areas. 12/7/2023 Not ready. 12/12/2023 12/28/22: plumbing passed, electrical passed, pending mechanical |
12/28/2023 | COT - HCME - Mechanical Above Ceiling | N/A | Not Ready | Scanner room above ceiling okay for hard lid. 12/13/2023 Ceiling grids for pre scan #103, work #104, control room #106. 12/20/2023 12/28/23: not ready |
COT - FCON - Fire - Final | N/A | Not Ready | subs need to call in their permit for the fire sprinklers. |
Completed Inspections
Permit Status: Complete
Permit Number: TC-COM-0523-01284
Permit Description: This project consists of a 879 sq.ft. remodel in an existing medical building for the installation of a Biograph Horizon equipment
Total Completed Inspections - 20