Inspection History Details
Inspection Status: Approved
Inspection History: PLUMBING - FINAL
Permit Number - T09CM01842
Inspection Description - PLUMBING - FINAL
Inspection Status - Approved
Date | Description | Inspector | Results | Comments |
---|---|---|---|---|
12/15/2009 | PLUMBING - FINAL | Approved | 1) GAS CLEARANCE REQUIRES INSPECTION OF THE GAS SYSTEM. I HAVE NOT INSPECTED IT AND NO ACCESS HAS BEEN PROVIDED 2) DELETION OF LAB ARRIER OK PER LETTER FROM ROBER PAGE ARCHITECT 3) LAB RESULTS OK 4) STERILIZATION PROCEDURE NEEDS TO BE DOCUMENTED ON THE COMPANIES LETTER HEAD |
|
12/14/2009 | PLUMBING - FINAL | Not ready for Inspection | IVRS - Inspection :* PROVIDE A STAMPED LETTER FROM YOUR ENGINEER OF RECORD DELETING THE LAV CARRIAGE SPECIFIED ON YOUR APPROVED PLAN, OR SEND YOUR REDLINED PLANS IN FOR REVISION AS PREVIOUSLY REQUESTED, OR INSTALL THE WORK PER YOUR LEGAL DOCUMENTS |
|
12/10/2009 | PLUMBING - FINAL | Denied | IVRS - Inspection : See detailed listing on site.| 999* 1) HOSE BIB VAC BREAKER 2) STAMPED LETTER FOR LAV CARRIER 3) CLEAN BT REPORT (PAPERWORK ON SITE INDICATES 1 OR MORE COLIFORM PRESENT) |
|
12/08/2009 | PLUMBING - FINAL | Not ready for Inspection | IVRS - Inspection :* 1) CANCELLED PER PHIL |
|
12/04/2009 | PLUMBING - FINAL | Not ready for Inspection | IVRS - Inspection :* 1) NOT READY PER MARCO |