Permit: T07CM03209

Parcel: 13811073B

Inspection Status: Approved

Inspection History: ELECTRIC - FINAL

Permit Number - T07CM03209

Inspection Description - ELECTRIC - FINAL

Inspection Status - Approved

Date Description Inspector Results Comments
04/22/2008 ELECTRIC - FINAL Approved IVRS - Inspection :*
04/21/2008 ELECTRIC - FINAL Denied IVRS - Inspection :
There is no access to the site.| 920*
NO ACCESS
BLD. LOCKED
04/18/2008 ELECTRIC - FINAL Denied IVRS - Inspection :
See detailed listing on site.| 999*
FLEX CABLE TO CHAIR TO COMPLY WITH 517.13 NEC
SEE DETAIL FOR CHAIR INSTALLATIONS REQUIRING HCF
04/15/2008 ELECTRIC - FINAL Denied IVRS - Inspection :
See detailed listing on site.| 999*
- CHAIRS WIRED IN NON-ACCEPTABLE WIRING METHOD
SEE MANUF. INSTRUCTIONS
USE APPROVED RACEWAY OR UNSPLICED CORD
LEAVE CHAIR ASSEMBLY AT TERMINATION OPEN FOR INSPECTION.
04/04/2008 ELECTRIC - FINAL Not ready for Inspection IVRS - Inspection :*
CHAIRS NOT INSTALLED YET