Permit: T07CM02853

Parcel: 133146610

Inspection Status: Approved

Inspection History: ELECTRIC - INTERMEDIATE

Date Description Inspector Results Comments
03/13/2008 ELECTRIC - INTERMEDIATE Approved IVRS - Inspection :*
11/02/2007 ELECTRIC - INTERMEDIATE Partial Approval IVRS - Inspection :*
OK WALLS & SOFFITS
11/01/2007 ELECTRIC - INTERMEDIATE Not ready for Inspection IVRS - Inspection :*
1. PROTECT MC CABLE AT TOP PLATES
2. SEAL UNUSED CONDUIT OPENING
3. EXISTING OPENINGS TO BE PER 517.13 - FLEX IS NOT A GROUND PATH PER 250
4. VDV TO BE SUPPORTED AS PER 725.6, 800.6, 300.11
5. BOX OVERFILLED
6. NEED TO SEE POWER FOR CHAIR LIGHTS
7. FIXTURES TO BE ATTACHED TO GRID 410.16
IF SCREWS 4 ON CORNERS OR 2 ON CENTER
09/18/2007 ELECTRIC - INTERMEDIATE Partial Approval IVRS - Inspection :*
OK U.G PER NOTES LAST INSP.
09/14/2007 ELECTRIC - INTERMEDIATE Partial Approval IVRS - Inspection :*
NOTE: EAST CHAIR CONDUIT ALMOST AT 360 DEG.

OK 3 POWER FLOOR BOXES

1. CONDUIT FOR VDV CABLING TO BE AS PER ARTICLE 300
2. CONDUIT FOR FUTURE NOT INSTALLED