Inspection History Details
Inspection Status: Approved
Inspection History: ELECTRIC - ABOVE CEILING
| Date | Description | Inspector | Results | Comments |
|---|---|---|---|---|
| 01/15/2003 | ELECTRIC - ABOVE CEILING | Approved | IVRS - Inspection 0000* | |
| 01/10/2003 | ELECTRIC - ABOVE CEILING | Partial Approval | IVRS - Inspection 0000*OK ABOVE CEILING. 1.) STILL NEED LETTER FOR M/C CABLE |
|
| 01/03/2003 | ELECTRIC - ABOVE CEILING | Partial Approval | IVRS - Inspection 0000*OKWALLS 1.) PLEASE SUPPORT OLD VDV CABLES UP OFF OF GRID. |
|
| 12/30/2002 | ELECTRIC - ABOVE CEILING | Partial Approval | IVRS - Inspection 0000*OK LIGHT FIXTURES EXCEPT NORTHWEST RESTROOM PLEASE NOTE THAN PLAN CALLS FOR WIRING TO BE IN 1/2" CONDUIT, USE OF MC MUST BE APPROVED BY ARCH. ALSO IN PATIENT CARE/EXAM ROOMS MUST COMPLY WITH ART 517-13 NEC. WIRING METHODS. |