Microfiche records prior to 2006 have not been completely digitized and may not be available yet on PRO. If you can not find what you are looking for please submit a records request.
Permit: T02CM00123
Permit Details
Apply Date:
01/09/2002
Applicant:
N/A
Description:
INSTALL:3 DENTAL CHAIRS
Architects:
- STEPHEN LEON MOOS
Contractors:
- SCHNEIDER BROTHERS CONSTRUCTION
Permit Reviews
Permit Number - T02CM00123
Permit Status: Withdrwn
Permit Description: INSTALL:3 DENTAL CHAIRS
Start Date | Submittal | Complete Date | Status |
---|---|---|---|
01/09/2002 | OTHER | 01/10/2002 | COMPLETED |
Outstanding Activity Conditions
Permit Number - T02CM00123
Permit Status: Withdrwn
Permit Description: INSTALL:3 DENTAL CHAIRS
Total Outstanding Activity Conditions - 1
Date | Name |
---|---|
01/09/2002 | PLANS EXPIRE IN 180 DAYS FROM DATE OF SUBMITTAL. PERMITS EXPIRE IN 180 DAYS FROM DATE OF ISSUE OR LAST INSPECTION. |
Outstanding Inspections
Permit Status: Withdrwn
Permit Number: T02CM00123
Permite Description: INSTALL:3 DENTAL CHAIRS
Total Outstanding Inspections - 6
Last Inspection Date | Description | Inspector | Results | Comments |
---|---|---|---|---|
BOILER - AIR TANK | N/A | Not Inspected | ||
01/24/2002 | ELECTRIC - UNDERGROUND | N/A | Denied | IVRS - Inspection 0245*NEED TO INSPECT CONDUIT RUN ABOVE CEILING ELECTRICAN TO MEET INSPECTOR TOMORROW AM PER CONVERSATION WITH KIETH S. |
01/24/2002 | ELECTRIC - INTERMEDIATE | N/A | Denied | IVRS - Inspection 0245* |
01/25/2002 | PLUMBING - INTERMEDIATE | N/A | Partial Approval | IVRS - Inspection 0000*1.) AIR PRESSURE TEST FOR WATER & VAC LINES OK .2) NOTE: MUST ANCHOR COPPER WATER LINE. |
01/28/2002 | ELECTRIC - ABOVE CEILING | N/A | Denied | IVRS - Inspection 0245*NOTES CONTR. USED EXISTING CIRCUITS DOES NOT COMPLY WITH REQUIREMENTS OF 517 NEC. BECAUSE HOME RUNS BAACK TO PANEL NOT USING METALIC RACEWAY. |
02/22/2002 | PLUMBING - FINAL | N/A | Denied | 1.)MUST PROVIDE LISTING FOR VACCUME PUMP ON WATER SUPPLY AIR BREAK AND VACCUME BREAKERS. |
Completed Inspections
Permit Status: Withdrwn
Permit Number: T02CM00123
Permit Description: INSTALL:3 DENTAL CHAIRS
Total Completed Inspections - 3
Date | Description | Inspector | Results | Comments |
---|---|---|---|---|
01/25/2002 | PLUMBING - PROGRESS | TRINIVJR | Approved | IVRS - Inspection 0000* |
02/22/2002 | BOILER FINAL | CWILLIAM | Approved | IVRS - Inspection 0000*UNFIRED PRESS. VESSELL. |
02/22/2002 | ELECTRIC - FINAL | DSANTACR | Approved | IVRS - Inspection 0000*CONTRACTOR TO RECALL WHEN ON SITE TO WALK THRU WITH INSPECTOR TO EXPOSE CEILING TO INSPECT CONDUIT ROUTING. |