Inspection Status: Approved
Inspection History: PLUMBING - MEDICAL GAS/TEST
| Date |
Description |
Inspector |
Results |
Comments |
| 05/07/2018 |
PLUMBING - MEDICAL GAS/TEST |
|
Approved |
GAS LINES: 1/2" - 1 1/2"; TESTED AT 150 PSI. OXYGEN, VACUUM, NITROUS OXIDE, WAGD, MEDICAL AIR CARBON DIOXIDE. |
| 01/10/2018 |
PLUMBING - MEDICAL GAS/TEST |
|
Partial Approval |
CARBON DIOXIDE LINE AT 1/2" OUTSIDE OF PROCEDURE ROOM #5. |
| 12/19/2017 |
PLUMBING - MEDICAL GAS/TEST |
|
Partial Approval |
CARBON DIOXIDE LINE AT 1/2" - UP TO JUNCTION OF EXPANXION JOIN. PHASE I IS COMPLETE. |
| 12/14/2017 |
PLUMBING - MEDICAL GAS/TEST |
|
Partial Approval |
INFORMATION ONLY - WILL SCHEDULE INSPECTION FOR PHASE I ON 18 DEC. |