Inspection History Details
Inspection Status: Approved
Inspection History: PLUMBING - PROGRESS
Permit Number - T09CM01776
Inspection Description - PLUMBING - PROGRESS
Inspection Status - Approved
Date | Description | Inspector | Results | Comments |
---|---|---|---|---|
08/25/2009 | PLUMBING - PROGRESS | Approved | IVRS - Inspection :* | |
08/18/2009 | PLUMBING - PROGRESS | Partial Approval | IVRS - Inspection :* 1) DV AND DA TESTING IN PLACE. LABELING ? 2) PROVIDE STAMPED LATER FROM ENGINEER GIVING OK FOR DV TRUNK LINE SIZE CHANGE |
|
08/17/2009 | PLUMBING - PROGRESS | Partial Approval | IVRS - Inspection :* 1) DETAIL AIR PIPING COMPLETE FROM COMPRESSOR ROOM THRU SYSTEM EXCEPT ROOM 109 DELETED 2) AIR TEST AT 150 PLUS CONTINUING ON DENTAL VACUUM PER APPROVED PLAN |
|
08/14/2009 | PLUMBING - PROGRESS | Partial Approval | IVRS - Inspection :* 1) INSTALLING DA AND DV, DA IS BEING FIT SYSTEM IS CHARGED WITH NITROGEN WHILE CONTINUING PIPING |