Permit: T09CM00133

Parcel: 11615088L

Inspection Status: Approved

Inspection History: PLUMBING - MEDICAL GAS/TEST

Permit Number - T09CM00133

Inspection Description - PLUMBING - MEDICAL GAS/TEST

Inspection Status - Approved

Date Description Inspector Results Comments
04/27/2009 PLUMBING - MEDICAL GAS/TEST Approved IVRS - Inspection :*
03/19/2009 PLUMBING - MEDICAL GAS/TEST Partial Approval IVRS - Inspection :*
TESTING TODAY
03/18/2009 PLUMBING - MEDICAL GAS/TEST Partial Approval IVRS - Inspection :*
1. INSTALLING LAST OF PIPE TODAY
2. RECALL 03045 AFTER TESTING AND CARD INITIALED
03/17/2009 PLUMBING - MEDICAL GAS/TEST Partial Approval IVRS - Inspection :*
MED-GAS PIPING AT STORAGE YARD OK
03/13/2009 PLUMBING - MEDICAL GAS/TEST Partial Approval IVRS - Inspection :*
ONE INSTALLER ON TERMINATIONS
03/12/2009 PLUMBING - MEDICAL GAS/TEST Not ready for Inspection IVRS - Inspection :*
03/10/2009 PLUMBING - MEDICAL GAS/TEST Partial Approval IVRS - Inspection :*
MEDICAL GAS DROPPED THRU ROOF AND CONNECTED TO CONTROL BOXES
03/05/2009 PLUMBING - MEDICAL GAS/TEST Not ready for Inspection IVRS - Inspection :*
1. CENTER LINE NOT ON SITE -- REQUESTED AM INSPECTION
2. NUMBER LEFT BY MIKE 473-1761 IS A NON-WORKING NUMBER
03/03/2009 PLUMBING - MEDICAL GAS/TEST Partial Approval IVRS - Inspection :*
1. GANG BOX OK
2. NITOGEN STORAGE OK
3. SUPPORTS OK
4. CONTINUING ROOF TOP PIPING
03/02/2009 PLUMBING - MEDICAL GAS/TEST Partial Approval IVRS - Inspection :*
1. PLEASE LEAVE A NAME AND CONTACT NUMBER WHEN SCHEDULING THESE INSPECTIONS, AS REQUESTED DURING THE PRE-CON MEETING. I DO NOT KNOW WHERE THE WORK AREA IS LOCATED AND EITHER WILL ANOTHER INSPECTOR FILLING IN FOR ME SHOULD I NOT BE AVAILABLE
2. ROOFTOP PIPING BEING INSTALLED OK
MED GAS BOX OK
NITROGEN TANK SECURE
SAFETY PRECAUTION BEING OBSERVED
02/26/2009 PLUMBING - MEDICAL GAS/TEST Partial Approval IVRS - Inspection :*
1. CERTIFIED PIPE MEDICAL GAS/MICHAEL D. SILVESTER
2. CLEAN AND BAGGED FITTING ON SITE
3. 02 ANALYZER ON SITE
4. TYPE K COPPER ON SITE 1 1/2