Permit: T08CM01129

Parcel: 13355231A

Inspection Status: Approved

Inspection History: ELECTRIC - FINAL

Permit Number - T08CM01129

Inspection Description - ELECTRIC - FINAL

Inspection Status - Approved

Date Description Inspector Results Comments
09/12/2008 ELECTRIC - FINAL Approved APPROVED PER SEALED MCGETRICK REPORT
09/11/2008 ELECTRIC - FINAL Partial Approval IVRS - Inspection :*
1. EQUIPMENT IS NOT ACCEPTABLE WITHOUT NRTL LISTING
2. THIS INSPECTOR DOES NOT HAVE AUTHORITY TO ACCEPT LETTER FROM ENGINEER.
3. NEED TO APPEAL TO BUILDING OFFICIAL JESSIE SANDERS FOR ACCEPTANCE OF UNLISTED X-RAY TABLE AND CHEST X-RAY GRID CABINET. JESSIE SANDERS @ 837-4900
09/08/2008 ELECTRIC - FINAL Denied IVRS - Inspection :
There is no access to the site.| 920*
UNIT LOCKED, NO ACCESS
09/04/2008 ELECTRIC - FINAL Partial Approval IVRS - Inspection :*
1. NO LISTING MARK ON X-RAY TABLE & GRID CABINET
PLEASE CHECK TO SEE IF EQUIPMENT IS LISTED, MAY NEED TO HAVE FIELD EVALUATION DONE OF LISTED EQUIPMENT INSTALLED. 14NN
2. ELECTRICAL INSTALLATION COMPLETE CONTRACTOR NOT RESPONSIBLE FOR X-RAY EQUIPMENT
08/28/2008 ELECTRIC - FINAL Partial Approval IVRS - Inspection :*
1. PREVIOUS INSPECTION ITEMS CORRECTED
2. RECALL AFTER X-RAY EQUIPMENT INSTALLED
08/27/2008 ELECTRIC - FINAL Partial Approval IVRS - Inspection :*
1. INSTALL EXTENSIONS ON RECEPTACLES & SWITCHES MORE THEN 1/4" BACK
2. MISSING EM LIGHT
3. WILL NEED TO SEE X-RAY EQUIPMENT WHEN INSTALLED
08/22/2008 ELECTRIC - FINAL Not ready for Inspection IVRS - Inspection :*
RECONNECT INSPECTION