Inspection History Details
Inspection Status: Approved
Inspection History: ELECTRIC - INTERMEDIATE
Permit Number - T08CM01037
Inspection Description - ELECTRIC - INTERMEDIATE
Inspection Status - Approved
| Date | Description | Inspector | Results | Comments |
|---|---|---|---|---|
| 03/13/2009 | ELECTRIC - INTERMEDIATE | Approved | IVRS - Inspection :* STILL NEED TO SEE T-BAR CEILING WALLS AND HARD LIDS OK |
|
| 03/11/2009 | ELECTRIC - INTERMEDIATE | Denied | IVRS - Inspection : See detailed listing on site.| 999* 1. PVC NOT ALLOWED 2. MC 1 1/4 FROM STUD 3. STRAPPING EVERY 5'-18" ACCORDING TO PRINT 4. EXIT NEEDED AT FRONT DOOR |
|
| 11/24/2008 | ELECTRIC - INTERMEDIATE | Partial Approval | IVRS - Inspection :* OK PIPE IN WALL |
|
| 11/17/2008 | ELECTRIC - INTERMEDIATE | Partial Approval | IVRS - Inspection :* 1. OK HARD LIDS PHASE 1 2. INSTALL HEALTHCARE CABLE TO RECEPT IN RM. 117 OK ALL OTHER WALLS PHASE 1 |
|
| 09/11/2008 | ELECTRIC - INTERMEDIATE | Partial Approval | IVRS - Inspection :* OK U.G RACEWAYS TO EXAM ROOMS -- PHASE 1 |