Date Description Inspector Results Comments
04/10/2009 ELECTRIC - FINAL Approved IVRS - Inspection :*
RECEIVED COPY OF FIELD EVALUATION
03/23/2009 ELECTRIC - FINAL Partial Approval IVRS - Inspection :*
ALL NEW IMAGING EQUIPMENT & ASSOCIATED SUPPORT EQUIPMENT TO BE LISTED - PLEASE VERIFY LISTING OF EQUIPMENT
02/19/2009 ELECTRIC - FINAL Partial Approval IVRS - Inspection :*
1. INSTALL PANEL DIRECTORIES
2. NEED PLAQUE @ SES STATING LOCATION OF BACK UP POWER LOCATIONS AND TYPES
3. PROVIDE TEST REPORT FOR NEW GEN. SET
4. INSTALL BONDING BUSHINGS OR RACEWAYS THROUGH CONCENTRICS WHEN 480V/277V
5. WHEN IS MEDICAL EQUIPMENT TO BE INSTALLED