Inspection History Details
Inspection Status: Approved
Inspection History: ELECTRIC - INTERMEDIATE
Permit Number - T07CM03987
Inspection Description - ELECTRIC - INTERMEDIATE
Inspection Status - Approved
| Date | Description | Inspector | Results | Comments |
|---|---|---|---|---|
| 03/18/2008 | ELECTRIC - INTERMEDIATE | Approved | IVRS - Inspection :* | |
| 03/13/2008 | ELECTRIC - INTERMEDIATE | Denied | IVRS - Inspection : See detailed listing on site.| 999* PARTIAL INSPECTION FOR X-RAY UNIT 1. WHITE CONDUCTOR NOT TO BE USED AS HOT 2. PROPERLY TERMINATE CONDUCTORS UNDER LUGS, BOND BUSHING AROUND REDUCING WASHERS 3. WIRING NOT PER PLAN |
|
| 01/25/2008 | ELECTRIC - INTERMEDIATE | Partial Approval | IVRS - Inspection :* OK WALLS |