Inspection History Details
Inspection Status: Approved
Inspection History: ELECTRIC - FINAL
| Date | Description | Inspector | Results | Comments |
|---|---|---|---|---|
| 02/06/2007 | ELECTRIC - FINAL | Approved | IVRS - Inspection :* | |
| 02/05/2007 | ELECTRIC - FINAL | Not ready for Inspection | IVRS - Inspection :* | |
| 01/05/2007 | ELECTRIC - FINAL | Partial Approval | IVRS - Inspection :* OK TEMP C OF O 1. NO EQUIPMENT TO BE USED IN PATIENT CARE AREAS WITHOUT INSPECTION FOR LISTING & LABELING 2. FINISH PREVIOUS ITEMS 4, 5, 6 |
|
| 01/03/2007 | ELECTRIC - FINAL | Partial Approval | IVRS - Inspection :* 1. FIRE ALARM PANEL PROTECT CABLES FROM SHARP METAL EDGES PROVIDE EARTH GROUND. 2. BOXES MORE THAN 1/4" FROM BOX TO SURFACE NEED APPROVED EXTENDER RING. 3. GROUND X-RAY EXAM LIGHTS 4. NEED TO SEE ALL MEDICAL EQUIPMENT NOTE: EQUIPMENT TO BE LISTED AND LABELED 110.3 5. EMERGENCY LIGHTING TO BE ON CIRCUIT FOR AREA SERVED 700.12 6. FINISH BRACE KITS 7. CHECK ABOVE CEILING AT BACK OF GUTTER NEXT TIME |
|
| 11/22/2006 | ELECTRIC - FINAL | Not ready for Inspection | IVRS - Inspection :* 1) WHEN WAS CONDUIT FOR SIGN INSPECTED ? 2) MEET WITH JEFF/CONTRACTOR TO DISCUSS SERVICE |