Permit: T05CM06322

Parcel: 12710229D


Addresses:

677 N WILMOT RD

Inspection Status: Approved

Inspection History: ELECTRIC - INTERMEDIATE

Date Description Inspector Results Comments
04/17/2006 ELECTRIC - INTERMEDIATE Approved IVRS - Inspection :*
04/12/2006 ELECTRIC - INTERMEDIATE Not ready for Inspection IVRS - Inspection :*
NO WORK PROGRESS
04/11/2006 ELECTRIC - INTERMEDIATE Partial Approval IVRS - Inspection :*
OK TEMP C OF O SUBJECT TO BUILDING INSPECTOR.
1. WILL NEED CITY STAMPED REVISIONS FOR FINAL.
04/10/2006 ELECTRIC - INTERMEDIATE Not ready for Inspection IVRS - Inspection :*
WAITING ON NRTL REP TO TEST EQUIPMENT
PLEASE CONTACT ME AT 909-4749 RICK
04/06/2006 ELECTRIC - INTERMEDIATE Not ready for Inspection IVRS - Inspection :*
PAPER WORK SUPPLIED IS NOT A FIELD EVALUATION REPORT OR LABELING AS SUCH. PLEASE CONTACT APPROPRIATE NRTL FOR TESTING OF ASSEMBLY.
03/31/2006 ELECTRIC - INTERMEDIATE Partial Approval IVRS - Inspection :*
03/28/2006 ELECTRIC - INTERMEDIATE Not ready for Inspection IVRS - Inspection :*
1. X-RAY EQUIPMENT NEEDS AN ASSEMBLY LISTING.
2. SUPPORT M/C CABLE AS PER CODES 12" & 6'
3. FINISH WALL WORK
4. PNAEL MORE THAN 1/4" FROM CAN TO SURFACE
5. FINISH INSTALLING COVERS ON WALL DUCTS
6. NEED CITY STAMP ON REVISIONS
7. CONDUCTORS AT 100A X-RAY DISCONNECT NOT AS PER PLANS.
8. NEED GROUNDING EQUIPMENT CONDUCTOR AT X-RAY DISCONNECT.
03/24/2006 ELECTRIC - INTERMEDIATE Not ready for Inspection IVRS - Inspection :*