Permit: T05CM05532

Inspection Status: Approved

Inspection History: PLUMBING - MEDICAL GAS/FINAL

Date Description Inspector Results Comments
08/22/2006 PLUMBING - MEDICAL GAS/FINAL Approved IVRS - Inspection :*
08/17/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
NOT A FINAL INSPECTION
RELOCATE MV
ALARMED REGULATOR REQUIRED
08/15/2006 PLUMBING - MEDICAL GAS/FINAL Not ready for Inspection IVRS - Inspection :*
06/05/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
1. LABELING OK
2. OK TO COVER ALL AREAS EXCEPT DOUBLE ROCKED M.R.I ROOM
3. CHECKED PAPERWORK
05/25/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
SOFT WATER PIPING ABOVE CEILING OK.
WORK ONGOING IN CORRIDOR 136.
CAPPED PIPE ON HANGERS IN CORRIDOR 111.
BLOW DOWN AND PURGE DOCUMENTATION CHECKED.
PROVIDE RED LINE DRAWINGS FOR CHANGES.
05/23/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
1. OK TO COVER WALLS - SUBJECT TO SPECIAL INSPECTOR.
2. CHECKED PAPER WORK
05/22/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
05/19/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
1. OK TO COVER ALL WALLS IN CORRIDOR #139
LABELS OK
2. SUBJECT TO SPECIAL INSPECTOR
CHECKED PAPERWORK - PLUMBER CAUGHT UP.
05/17/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
1. DISCUSSED THE IMPORTANCE OF DAILY PAPERWORK WITH INSTALLER.
05/15/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
N. WALL ONLY ROOMS EXAM 144, 146, 164 LABELING & SUPPORTS OK
FROM ZONES IN CORRIDOR 139
05/12/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
1. CHECKED PAPERWORK
2. PROGRESS GOOD
05/10/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
1. UNABLE TO LOCATE INSTALLER IN ORDER TO CHECK PAPERWORK.
2. IMPORTANT!! ALL MEDICAL GAS PIPING TO REMAIN CAPPED OR PLUGGED AT ALL TIMES. WILL NOT ACCEPT ANYTHING LESS.
05/09/2006 PLUMBING - MEDICAL GAS/FINAL Not ready for Inspection IVRS - Inspection :*
NO ONE ON SITE TO PROVIDE ACCESS TO WORK LOG.
05/07/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
CHECKED ON PAPERWORK
PROGRESS OK
05/04/2006 PLUMBING - MEDICAL GAS/FINAL Partial Approval IVRS - Inspection :*
1. MED GAS SET UP - GANG BOX OK
STORAGE OK
04/26/2006 PLUMBING - MEDICAL GAS/FINAL Not ready for Inspection IVRS - Inspection :*
1. ADD 03045 & 03089 TO PERMIT
2. MED. GAS PRE CON
3. M&M PLUMBING
CITY OF TUCSON
COSTAS MEDICAL
RIO WEST
3. CALL 03045 WHEN PLUMBER BEGINS INSTALLING.