Permit: T05CM05036

Parcel: 11615088L

Inspection Status: Approved

Inspection History: ELECTRIC - INTERMEDIATE

Permit Number - T05CM05036

Inspection Description - ELECTRIC - INTERMEDIATE

Inspection Status - Approved

Date Description Inspector Results Comments
02/10/2006 ELECTRIC - INTERMEDIATE Approved IVRS - Inspection :*
01/27/2006 ELECTRIC - INTERMEDIATE Denied IVRS - Inspection :
See detailed listing on site.| 999*
1. NEED TO PROVIDE ACCESS TO SUB-PANEL RACEWAYS
2. GROUNDING NOT CORRECT @ X-FORMER NEC 250.30
3. NL5-8 CKT IS DEAD
4. INSTALL GFI BY SINK CORRECTLY
5. COMPLETE WIRING FOR DISCONNECT
6. MAKE SURE ALL WIRING
OK TO OCCUPY PRE-POST AREA FOR TEMPORARY USE
01/23/2006 ELECTRIC - INTERMEDIATE Not ready for Inspection IVRS - Inspection :*
01/11/2006 ELECTRIC - INTERMEDIATE Partial Approval IVRS - Inspection :*OK TO COVER ELECTRICAL IN WALLS
NOTE: PROVIDE SUPPORT FOR BOXES AS NEEDED
12/01/2005 ELECTRIC - INTERMEDIATE Partial Approval IVRS - Inspection :* OK TO COVER ELECTRICAL IN PRE POST HOLDING WALLS AFTER ALL BOX'S ARE PROPERLY SUPPORTED
OK TO COVER ELECTRICAL CONDUIT UNDERGROUND FOR CATH LAB WHEN FINISHED