Permit: T05CM03639

Parcel: 11615190K

Inspection Status: Approved

Inspection History: ELECTRIC - FINAL

Permit Number - T05CM03639

Inspection Description - ELECTRIC - FINAL

Inspection Status - Approved

Date Description Inspector Results Comments
12/09/2005 ELECTRIC - FINAL Approved IVRS - Inspection :*
OK ELECTRIC FINAL
12/08/2005 ELECTRIC - FINAL Denied IVRS - Inspection :
The wiring is not complete or does not comply with code.| 245*
INSTALL GFI PROTECTION WHERE PLAN REQUIRES.
1. IDENTIFY PANEL AND CORRECTLY MARK PANEL DIRECTORY'S
2. TIGHTEN CONNECTORS @ OUT FOR GAS SLAB (LIGHT ON LOAD SIDE GFI
3. APPROVED PLANS CALL FOR WIRING TO COMPLY WITH ARTICLE 517 BUT APPEARS TO BE SEVERAL VIOLATIONS, ESPECIALLY IN PATIENT CARE AREAS - PLEASE CHECK