Permit: T05CM03250

Parcel: 11010298A


Addresses:

3150 N SWAN RD

Inspection Status: Approved

Inspection History: PLUMBING - ABOVE CEILING

Date Description Inspector Results Comments
09/07/2006 PLUMBING - ABOVE CEILING Approved IVRS - Inspection :*
08/09/2006 PLUMBING - ABOVE CEILING Denied IVRS - Inspection :
See detailed listing on site.| 999*
PIPING ON UNISTRUT TRAPEZE ALONG WEST WALL IS MISSING SOME UNISTRUT CLAMPS.
MED GAS PIPING AT TEST GAUGE LOCATION NEEDS SUPPORT.
PROVIDE PIPE INSULATION SPECIFICATIONS FOR THIS JOB.
06/05/2006 PLUMBING - ABOVE CEILING Partial Approval IVRS - Inspection :*