Permit: T05CM02761

Parcel: 13421093A

Inspection Status: Approved

Inspection History: MEDICAL - GAS/TEST INTERMED

Date Description Inspector Results Comments
09/21/2005 MEDICAL - GAS/TEST INTERMED Approved IVRS - Inspection :*
1) INITIAL PRESSURE TEST COMPLETE
2) RESULTS ON SITE
3) PIPE HUNG AND LABELED OK
4) OK TO COVER ALL AREAS
09/20/2005 MEDICAL - GAS/TEST INTERMED Not ready for Inspection IVRS - Inspection :*
PLEASE SEE:
CALLED IN LAST FRIDAY. INSP NOT DONE.

5.3.10.10.6.1 PIPE SUPPORT NFPA 99C
ALSO IS-9 94WPC
HANGER AT EACH CHANGE OF DIRECTION, OR ELEVATION AND AT THE END OF BRANCHES THEN NO MORE THAN FOUR FEET BETWEEN HANGERS
NOTE: PLUMBERS TAPE NOT APPROVED STRAP