Inspection Status: Approved

Inspection History: MECHANICAL - ABOVE CEILING

Date Description Inspector Results Comments
10/28/2005 MECHANICAL - ABOVE CEILING Approved IVRS - Inspection :*
10/20/2005 MECHANICAL - ABOVE CEILING Partial Approval IVRS - Inspection :* OPER. RM 135/133 PROGRESSING. OK TO COVER BALANCE OF GRID.
10/11/2005 MECHANICAL - ABOVE CEILING Partial Approval OK TO COVER WEST SIDE GRID. TEST # 12. NIT OXIDE, NITROGEN , WASGD, 60 PSI, OXYG & VAC. MED. GAS PSI. MED GAS ROM # 125 MECH ROOM # 124- Inspection :*
09/16/2005 MECHANICAL - ABOVE CEILING Partial Approval IVRS - Inspection :*
FIRE DAMPERS MUST BE ACCESSABLE!
08/29/2005 MECHANICAL - ABOVE CEILING Partial Approval IVRS - Inspection :*
1) ITEM #4005 NOT APPLICABLE PER PAPERWORK TURNED IN BY INSPECTOR C. WILLAMS
2) MED GAS PRESSURE TEST #5/PRE-OP ROOMS #148: 1, 2, 3, 4, 5 AND 6
3) 02 MED AIR AND MED VACCUM
08/25/2005 MECHANICAL - ABOVE CEILING Partial Approval IVRS - Inspection :*
1) TEST PRESSURE #3 RECOVERY ROOMS A AND C-LOW PRESSURE TEST 60 PSI, FULL LOOP AND STATIONS
2) TEST PRESSURE #4 TREATMENT ROOMS 02 AND VACCUM LINE 160 PSI