Inspection History Details
Inspection Status: Approved
Inspection History: ELECTRIC - INTERMEDIATE
Permit Number - T05CM01498
Inspection Description - ELECTRIC - INTERMEDIATE
Inspection Status - Approved
| Date | Description | Inspector | Results | Comments |
|---|---|---|---|---|
| 07/05/2005 | ELECTRIC - INTERMEDIATE | Approved | IVRS - Inspection :* | |
| 06/08/2005 | ELECTRIC - INTERMEDIATE | Partial Approval | IVRS - Inspection :* | |
| 05/27/2005 | ELECTRIC - INTERMEDIATE | Not ready for Inspection | IVRS - Inspection :* / 1) SUPPORT UP OLD NM CABLES AND VDV CABLING TUBE ALSO SUPPORTED OR REMOVED OR TAGGED PER 725.2, 725.6, 800.2 AND 800.6 - 2) NM CABLES ABOVE PANEL TO BE SUPPORTED NOT OFF OF WIRING METHODS 334.30 ALSO DO NOT BUNDLE CABLES. 3) RELOCATED LIGHTING NOT INSTALLED AT THIS TIME. 4) AREAS OF CONSTRUCTION STILL IN USE. | |
| 05/16/2005 | ELECTRIC - INTERMEDIATE | Partial Approval | IVRS - Inspection :* OK WALLS FOR TREATMENT ROOMS. 1) FINISH PANEL WALLS. 2) AREA'S OF CONSTRUCTION ARE BEING OCCUPIED WITHOUT C OF O DURING CONSTRUCTION. |