Microfiche records prior to 2006 have not been completely digitized and may not be available yet on PRO. If you can not find what you are looking for please submit a records request.
Inspection History Details
Inspection Status: Approved
Inspection History: MEDICAL - GAS/TEST INTERMED
Permit Number - T03CM05171
Inspection Description - MEDICAL - GAS/TEST INTERMED
Inspection Status - Approved
Date | Description | Inspector | Results | Comments |
---|---|---|---|---|
02/10/2004 | MEDICAL - GAS/TEST INTERMED | Approved | 1) NO INSPECTIONS MADE AT START OF JOB. ALL INSPECTIONS, PAPERWORK WILL BE BY ENGINEER OF RECORD, CONTRACTOR AND THIRD PARTY CERTIFYING SYSTEM | |
01/22/2004 | MEDICAL - GAS/TEST INTERMED | Cancelled | IVRS - Ri canceled | |
01/14/2004 | MEDICAL - GAS/TEST INTERMED | Partial Approval | IVRS - Inspection 0000*/ PARTIAL INSPECTION ON COMPRESSED AIR LINE. CONFIRMED PLUGGED PIPE, BAGGED FITTINGS, NITRGEN. OK TO PROCEED. | |
01/02/2004 | MEDICAL - GAS/TEST INTERMED | Partial Approval | IVRS - Inspection 0000*/ ENGINEER OF RECORD HAS CALLED WITH CONCERNS - PLEASE CALL ME SO MEDICAL GAS ISSUES CAN BE DISCUSSED. THANKS. | |
01/05/2003 | MEDICAL - GAS/TEST INTERMED | Denied | RESPONDED TO A COMPLAINT FROM AL NICHOLS, THE ENGINEER OF RECORD FOR THIS PROJECT, ABOUT THE CONTRACTOR NOT INSTALLING MED. GAS PIPING PER PLAN. SPOKE WITH CHUCK FROM CAYLOR CONSTRUCTION AND INFORMED HIM ANY DEVIATION FROM THE PLANS MUST BE REVISED ON PLANS AND SENT BACK FOR PLAN REVIEW. |