Inspection History Details
Inspection Status: Approved
Inspection History: ELECTRIC - INTERMEDIATE
Permit Number - T01CM00034
Inspection Description - ELECTRIC - INTERMEDIATE
Inspection Status - Approved
Date | Description | Inspector | Results | Comments |
---|---|---|---|---|
10/19/2001 | ELECTRIC - INTERMEDIATE | Approved | ||
06/22/2001 | ELECTRIC - INTERMEDIATE | Partial Approval | IVRS - Inspection 0000*OK WALLS SUBJECT TO LETTER OF APPROVAL FOR USE OF M/C CABLE | |
06/13/2001 | ELECTRIC - INTERMEDIATE | Denied | IVRS - Inspection 0999*1)M/C CABLE HAS TO BE LISTED FOR HOSPITAL/PATIENT CARE USE ART 517. 2)NEED REDUNDAT GROUND IN PATIENT CARE. 3)LOOSE FITTINGS AND LOCKOUT SCONDUIT AND M/C CABLE.4)PROTECT M/C CABLE WHERE CROSS OPEN STUD.5)FILL UNUSED OPENINGS IN PANEL. 6)MISSING PLASTER RING.7)NEED LETTER FROM ARCH. AUTHORIZING USE OF M/C CABLE.8)NEED BACK BRACING AS PER SPEC'S.#10-F |