Case Details

Status:

COMPLIAN

Type:

FIRE AT COMMERCIAL BUILDING

Entered Date:

07/03/2017

Applicant:

ITE LTD

Description:

COMMUNICATION CALLED THAT T.F.D. ON SITE WITH AN X-RAY MACHINE FIRE, T1747723

Outstanding Inspections

Case Number: T17DV02871
Case Status: COMPLIAN
Case Description: FIRE AT COMMERCIAL BUILDING
Total Outstanding Inspections - 7
End Date Description Inspector Results Comments
07/04/2017 EMERGENCY CALL OUT VIDAL Item closed
07/04/2017 NPO(EL)-MAIN/PANEL/SUB/DISCONT VIDAL Item closed
07/04/2017 NPO(EL)-WIRING/SWITCH/OUTLT/FX VIDAL Item closed
07/04/2017 NPO(HC)-COOLING VIDAL Item closed
07/04/2017 NPO - INTERIOR TRASH/STORAGE VIDAL Item closed
07/04/2017 NPO - OTHER VIDAL Item closed
07/04/2017 NPO - INT. BUILDING MAINTENANC VIDAL Item closed

Completed Inspections

Case Status: COMPLIAN
Case Number: T17DV02871
Case Description: FIRE AT COMMERCIAL BUILDING
Total Completed Inspections - 2
End Date Description Inspector Results Comments
07/03/2017 INSP - INITIAL CODE ENFORCMENT VIDAL Approved Received call from communication that T.F.D. on site with an X-ray machine fire. Arrived and the units were still on site. The fire was in a control box and the transformer was smoking. The breakers to the X-ray machine and the control panel have been turned off on the interior of the building. Did turn off the disconnect and placed a City of Tucson lock for the X-ray machine only. All other electrical was left on at this time. The office has already call the tech's for the X-ray machine and will come to the site to check the X-ray machine. Dry powder was used on the equipment and is through out the complex. At this time the office will be closed until the office has been cleaned and the office can re-open once the cleaning crew has given the OK that the office is cleaned. Did do the posting on the front door of Do Not Enter. Place caution tape around the front doors due to the dust was still in the air and left the doors opened for fresh air in the building. Will check once the office calls that that the office is cleaned. The administrator signed the paperwork. RV.
07/04/2017 INSP - FOLLOW UP CODE ENFORCE VIDAL Approved Communication called and the adminstrator called to get clearance on opening on Wednesday. The X-ray tech's were on site and needed the electrical turned back on to check the system once the repairs have been made. The transformer burned and was replaced with new. The administrator has a letter on the office being cleaned and able to open for business. The X-ray machine will operable by the end of the day. At this time the administrator was advised to call the health department to check if they need to have the area inspected and if not are able to open on Wednesday. The lock on the disconnect was removed and the electrical was left off until the tech's turn it back on. At this time the case will be closed. Violation Abated, Close Case. RV.

Completed Activity Special Conditions

Case Description - FIRE AT COMMERCIAL BUILDING
Case Number - T17DV02871
Case Status: COMPLIAN
Total Completed Case Activity Conditions: 2
Date Name
07/25/2017 NOTICE OF VIOLATION ISSUED
07/25/2017 VIOLATION CLOSED